Sending a Switching Command from the Host Device
This method is convenient when the unit has already been mounted and the switch on the bottom can-
not be repositioned to "3."
During operation in the Normal Operation mode, a command is sent from the host device to switch to
the Setting mode.
1.
Send a subsystem command (S18F13 ChangeState CPVAL1 = "PS") from the host device.
page 70.
Refer to page
CPVAL1="PS" is an expansion designation unique to V700-L21 and does not conform to SEMI standards.
The system is automatically restarted and the mode switches to the Setting mode.
The operation indicators react as shown below.
OPERATING
ALARMS
Start the terminal software
Use the host device's terminal software for the setting.
The commands and communications conditions in the setting mode are unique to OMRON. They do not conform to the
SEMI standards. For the terminal software, use Hyper Terminal, which is standard with Windows, or a similar program.
The communications conditions for communication between the host device and CIDRW Controller
are fixed. Make the following settings using the terminal software.
Item
Baud rate
Data length
Parity
Stop bit
Communications control None
Send code
Display
BUSY
ERROR
9600 bps
8 bits
EVEN
1
At the end of a line (when [ENTER] is input), the "line feed" characters ([LF]) are appended.
Local echo
Preparing for Communications
Setting
SECTION 3
CIDRW System
45
User's Manual