PRESYS
Instruments
Another way of performing automatic documented calibrations is by means of
ISOPLAN Calibration Software for PC/Windows
communication to connect the computer to the dry block. With ISOPLAN it is possible to
register sensors and instruments of a factory, generate work orders, create and print
calibrations certificates and reports, that is, it brings all advantages of computer data
management to the calibration environment.
T-1200PH have also many other features, such as:
The electric signal calibrator is independent from the dry block function.
Internal buzzer beeps when the temperature reaches the desired value.
Keypad that eases the operation and configuration of the calibrator.
Graphic liquid crystal display to present big-size digit numbers.
Thermo-element reading scaled to ITS-90 or IPTS-68.
Internal regulated 24 vdc power supply for 2-wire transmitters.
Internal rechargeable battery and battery charger included in the electric signal
calibrator.
Independent circuitry for overtemperature protection and safety.
Insert to choose, carrying case, strap and test leads included. If the insert is not
specified, it will be provided the insert type BP06.
1.1 - Technical Specifications
Operating Range
Power Supply
Resolution
Accuracy
Stability
Power Consumption
Heating Rate
Cooling Rate
Calibration Volume
Standard Insert coming with
the furnace
Homogeneity
Dimensions (H,W,D)
Weight
Warranty
TM
, which uses RS-232 or RS-485 serial
50 ºC (122 °F) to 1200 °C (2192 °F)
110 Vac or 220 Vac 50/60Hz, according to order code.
45 minutes (100 ºC to 1200 ºC)
5 h (1200 ºC to 200 ºC)
34 mm / 130 mm depth
2 holes of 6 mm and 2 holes
of ¼" x 80 mm depth
215 x 390 x 310mm
T-1200PH
0.1 ºC or 0.1 ºF
3.0 °C
0.2 ºC
2300 W
0.2 °C
10,3 kg
1 year
T-1200PH