Presys T-1200PH Technical Manual page 27

Dry block calibrator
Table of Contents

Advertisement

PRESYS
 Instruments
superthermometer (ST). Now, go back to the option Calibration/ Probe / ºC and record
the values indicated by the superthermometer.
For T-1200PH:
Setpoint of the
temperature
generated (ºC)
150
350
600
750
850
1000
1100
Initial value to
record (ºC)
indication
150.0
350.0
600.0
750.0
850.0
1000.0
1100.0
1087.11
ST
New value to
record
149.96
150.0
349.93
349.9
598.03
598.0
745.32
745.3
843.13
843.1
990.45
990.4
1087.1
T-1200PH
New
indication
key
of the ST
150.01
key 1
349.99
key 2
600.02
key 3
749.99
key 4
850.03
key 5
999.97
key 6
1100.05
key 7
Page 21