TU: Menu that configures the PID control parameters for the heating side. More
details in section 6 - PID Control Parameters.
K: Proportional gain
I: Integral gain
D: Derivative gain
FACT: Restore the control parameters to the factory values.
PRG: Menu that programs the calibrator.
DEC_IN: Selects the number of decimals of the thermo-element reading.
The default value depends on the input signal.
DEC_PRB: Selects the number of decimals of the block temperature and
the setpoint value. The default number is 2.
SETPOINT: Enables the calibrator Programmable Operating Mode, and
allows the configuration of programmed values. The selected program is
indicated by the selection arrow. Choose one of the 6 temperature programs or
NO to disable this Programmable Mode.
Select any of the 6 programs and confirm with the key ENTER. Following, it is
shown the menu of temperature setpoint configuration.
Change the configuration to steps (STEPS) of 10%, 20%, 25%, VARIABLE or
press C/CE to maintain the configuration already stored in memory. The temperature
range of the program must be configured through the values in SETPOINT HIGH and
SETPOINT LOW in case of 10%, 20% or 25% fixed steps of the range. The option
VARIABLE allows the user to define from 2 to 11 values of temperature setpoint, not
necessarily in ascending order.