4 - Safety Instructions

 If the calibrator is turned on, do not leave the room without an identification or warning
about the high temperature hazard.
 Before turning the calibrator off, return the block temperature to values close to the
ambient temperature.
 Never remove the insert from the dry block or the thermo-elements from the insert,
while they are in temperatures far from the ambient. Wait until they reach the ambient
temperature.
 Never transport the dry block with the metallic insert inside it, as the metallic insert can
hit the ceramic tube damaging it permanently.

5 - Calibrator Warning Messages

Warning
RAM ERROR
READ MANUAL
EEPROM ERROR
READ MANUAL
LOW BATTERY
UNDER / OVER
????.??ºC
Meaning
Problem in RAM
memory
Problem in EEPROM
memory
Level of battery voltage is
low
Input signal out of
specifications or scaling
range
Input sensor is open
Procedure
Turn the calibrator off and
on. If the error persists,
send the instrument to the
factory
Same as the previous item
Check the T-1200PH
calibrator power supply
See item 1.1.1 on Input
Specifications
Check input connections
and sensor