1 - Introduction

T-1200PH dry block calibrators control temperature of an insert in order to
calibrate thermocouples, thermoresistances, glass thermometers, thermoswitches etc.
Besides providing high accuracy temperature values, they also allow the measurement of
signals generated by thermo-elements like thermocouples, thermoresistances and
thermoswitches, which are being calibrated. This is possible due to an embedded
calibrator specific for these types of signal, including 4-20 mA. Thus, they incorporate the
functions of dry block, standard thermometer and calibrator for RTD, TC and also mA.
T-1200PH calibrators model generate temperatures from 50 ºC (122 ºF) to 1200
ºC (2192ºF).
Present input for thermocouples, thermoresistances, thermoswitches. Besides
generating temperature, they measure the signal from the sensor being
calibrated.
Carry out completely automatic calibrations with or without the use of a computer.
Accuracy to 3 ºC of reading, stability of 0.2 ºC and resolution of 0.1 ºC.
Documenting
Calibration Software.
Portable, compact and, provide interchangeable inserts.
They present a wide variety of programming resources, allowing the performance
of automatic calibrations of thermocouples, thermoresistances, thermoswitches. In this
case, the sensor is placed in the insert and their electrical terminals are connected to the
embedded calibrator. The operator defines the calibration points and the number of
repetitions, then the process is started and all the sequence is automatically
accomplished.
capabilities:
communication
T-1200PH
with
computer
and
T-1200PH
ISOPLAN
Page 1