Safety Instructions; Calibrator Warning Messages - Presys T-350P Technical Manual

Dry block calibrators
Table of Contents

Advertisement

PRESYS
 Instruments

4 - Safety Instructions

 If the calibrator is turned on, do not leave the room without an identification or warning
about the high temperature hazard.
 Before turning the calibrator off, return the block temperature to values close to the
ambient temperature.
 Never remove the insert from the dry block or the thermo-elements from the insert,
while they are in temperatures far from the ambient. Wait until they reach the ambient
temperature so that the heterogeneous cooling of the parts do not jam each other. If,
by chance, it happens a jamming situation, refer to item 8.2 - Instructions for Insert
Jamming to proceed properly.
 Never interchange a T-650P insert to an insert from a different temperature range
calibrator (like T-350P), because in spite of having the same diameter, it is made of a
different material not resistant to high temperature.

5 - Calibrator Warning Messages

Warning
RAM ERROR
READ MANUAL
EEPROM ERROR
READ MANUAL
LOW BATTERY
UNDER / OVER
????.??ºC
Page 16
Meaning
Problem in RAM
memory
Problem in EEPROM
memory
Level of battery voltage is
low
Input signal out of
specifications or scaling
range
Input sensor is open
T-350 P / T-650 P
Procedure
Turn the calibrator off and
on. If the error persists,
send the instrument to the
factory
Same as the previous item
Check the T-350P and
T-650P calibrator power
supply
See item 1.1.1 on Input
Specifications
Check input connections
and sensor