Changing Filters

Disposable
 Prior to changing the filters, the water level in the tank must be no higher than the
level of the spray bar.
 Filters should be changed periodically depending on the usage of the equipment and
the contaminants being removed. If water is not flowing more than a trickle from the
spray bar holes, it is time to replace your filters.
 Individually unscrew counterclockwise the two (2) large blue filter cartridge holders.
The cartridges will be full of water.
 Pour the water out of the cartridges and remove the filters. Be careful not to lose the
rubber O-ring gasket located on the inside of the filter cartridge. Absence of this
gasket will cause the filter(s) to leak.
 Replace with new filter(s) and re-install the cartridges.
 The 20 Micron Polycarbonate Filter is the primary filter and is located on the right
side of the filtration unit.
 The 10 Micron Polycarbonate Filter is the secondary filter and is located on the left
side of the filtration unit.