Changing Filters

Disposable
· Prior to changing the filters, the water level in the tank must be no higher than the
level of the spray bar.
· Filters should be changed periodically depending on the usage of the equipment and
the contaminants being removed. If water is not flowing more than a trickle from the
spray bar holes, it is time to replace your filters.
· Individually unscrew counterclockwise the two (2) large blue filter cartridge holders.
The cartridges will be full of water.
· Pour the water out of the cartridges and remove the filters. Be careful not to lose the
rubber O-ring gasket located on the inside of the filter cartridge. Absence of this
gasket will cause the filter(s) to leak.
· Replace with new filter(s) and re-install the cartridges.
· The 50 Micron Polycarbonate Filter is the primary filter and is located on the right
side of the filtration unit.
· The Carbon Filter is the secondary oil filter and is located on the left side of the
filtration system. The Carbon Filter removes oil and grease and will require
changing more often than the 50 Micron Polycarbonate Filter.
For Fire, Mold, Water Restoration, the primary filter is a 20 micron and the
secondary is a 10 micron.
Reusable Stainless Steel
· Clean the outside of the filters in a sink.
· Place each filter into individual Ziplock bags.
· Fill each bag with hot water and add soap from the bath.
· Seal bag and place into ultrasonic tank and operate the ultrasound for 15 minutes.
· Dump the fluid from the bag into the sink, NOT the ultrasonic tank.
· Repeat the process until no dirt is extracted from the filters.
·
Replace the filters into the blue canisters.