The Pump is not operational.
1. Turn on the Pump and listen for the sound of the motor spinning.
2. Note: Pump motor is located on the left end of the machine under the blue
canisters.
3. Put your hand near the back of the pump motor and feel for any air flow.
4. If you feel air flow, or feel motor vibration or hear the motor humming, follow
steps a-f.
a. Unscrew each blue canister and remove the filters completely.
b. Place the filters in a sink or away from the machine and keep track of
orientation.
c. Note: Make sure to not lose black O-ring from blue canister or leaks will
occur.
d. Screw blue canisters back on.
e. Turn on Pump and wait several minutes to see if water moves through
spray bar.
f. If you have water flow, the filters are clogged and must be replaced.
5. If you are using stainless steel reusable filters, follow 5a - 5g.
a. Clean outside of filters in a sink.
b. Place each filter into individual Ziplock bags.
c. Fill each bag with hot water and add soap.
d. Seal bag and place into ultrasonic tank and operate ultrasound for 15
minutes.
e. Dump the fluid from the bag into a sink – not into the ultrasonic tank.
f. Repeat this process until no dirt is extracted from the filters.
g. Replace filters into the blue canisters.
6. If you do not have water flow without filters and the motor is operational, follow
6a - 6c.
a. Drain tank and remove the pump head volute from motor.
b. Inspect the impellor and shaft to see where the damage is.
c. Replacing the Pump Head less Motor is the likely solution.