Measurement
Cuff pressure 0-300 mmHg,
Range
Systolic 50-250 mmHg,
Diastolic 30-200 mmHg,
Pulse 40-180 beats/minute
Display Accur-
Systolic ± 3 mmHg,
acy
Diastolic ± 3 mmHg,
Pulse ± 5% of the value shown
Measurement
Max. permissible standard
inaccuracy
deviation according to clinical
testing: Systolic 8 mmHg /
Diastolic 8 mmHg
Memory
4 x 30 memory spaces
Dimensions
5.5 in L x 4.0 in W x 2.1 in H
(139 mm L x 103 mm W x 54
mm H)
Weight
9.4 oz (267 g) without cuff
Cuff Size
8.7 in – 17.3 in (22 – 44 cm)
Operating
Range
Storage Range
Power Supply
Battery Life
Accessories
Classification
24
41° F – 104° F (5° C – 40° C),
15-93% relative humidity (non-
condensing)
-13° F – 158° F (-25° C – 70° C),
< 93% relative humidity
4 x 1.5V
AA batteries
Approx. 200 measurements
(depending on levels of blood
pressure and pump pressure/
Bluetooth
transfer)
®
Instructions for use, 4 x 1.5V
AA batteries, storage pouch
Internal supply, IP21, no AP or
APG, continuous operation,
application part type BF