5.
If the soft front panel is not already running, click Run/Stop.
6.
Add oscilloscope measurements to the channel by selecting Add/Remove in the
measurement table of the oscilloscope.
For more information on the measurements available for oscilloscopes in InstrumentStudio,
refer to the InstrumentStudio Manual at ni.com/manuals.

Making a Measurement with LabVIEW

1.
Launch LabVIEW.
Select Help»Find Examples.
2.
3.
Open the example VI that you want to use by selecting Hardware Input and Output»
Modular Instruments»NI-SCOPE (High-Speed Digitizers).
Tip
found under Hardware Input and Output»Modular Instruments»NI-
SCOPE (High-Speed Digitizers)»Demos»niScope EX Quick Start.vi.
4.
Follow any setup instructions in the VI and specify any desired settings.
Click Run to run the example program.
5.

PXIe-5163 Compensating Passive Probes

Compensating passive probes increases the accuracy of your measurements by matching the
capacitance of your probe to the capacitance of a particular oscilloscope input channel.
Before beginning this procedure, complete the following:
Set up your oscilloscope in a compatible chassis;
Install a version of NI-SCOPE compatible with your oscilloscope on your system; and
Configure a panel and layout in InstrumentStudio to include your oscilloscope.
Note
InstrumentStudio Manual at ni.com/manuals.
The following equipment is required for this procedure:
If using the Oscilloscope Probe
Compensator
If connecting the probe directly to
the oscilloscope PFI
The PXIe-5163 can output a 1 kHz square wave that you can use to compensate passive
probes.
To compensate a passive probe, complete the following steps:
1.
Connect the BNC end of the probe to an input channel of your oscilloscope.
14 | ni.com | PXIe-5163 Getting Started Guide
If you are not sure which example to run, use the Quick Start VI, which is
For more information on using InstrumentStudio, refer to the
Oscilloscope Probe Compensator (NI part number
786983-01)
A 50 Ω SMB (f)-to-BNC (f) cable, such as the
SMB-100 (NI part number 781449-01)
A probe tip–to-BNC (m) adapter