No.
11
12
13
1.1.1.2
Compact version
No.
1
2
3
4
5
6
7
8
9
10
|
Quick reference guide
Designation
Anesthetic gas receiving system
External O
flowmeter (option)
2
Rail
Designation
Screen
Plug-in connectors for vaporizers Enable the connection of up to 2 vaporizers.
Backup manual switch (behind
the flap)
Folding table extension (option)
Work surface
Viewing window for piston venti-
lator
Lockable drawer
Trolley with castors and cable
deflectors
CO
absorber
2
Anesthetic gas receiving system
Atlan SW 2.0n
Description
Used for scavenging excess anesthetic gas and
breathing gas. Used to reduce the anesthetic
gas concentration released into the environment
by the anesthesia machine and to scavenge the
sample gases from an external anesthetic gas
monitor.
Supplies oxygen for O
2
Used for fastening additional components.
Description
Enables user inputs by touchscreen and rotary
knob.
Used to switch to backup manual mode in the
event of malfunctions of the device.
Used for storage.
Used for storage.
Allows visual checking of the movement of the
piston ventilator.
Provides additional storage space.
Used for moving the device. The central brake
(option) locks the two front castors.
Absorbs CO
from the patient's expiratory gas.
2
Used for scavenging excess anesthetic gas and
breathing gas. Used to reduce the anesthetic
gas concentration released into the environment
by the anesthesia machine and to scavenge the
sample gases from an external anesthetic gas
monitor.
Overview
insufflation.
5