8 Configuration and Setup

8.1

Signal Conditioning

Emissivity and Transmittance (Spectral mode)
Besides the emissivity correction (mentioned in Chapter 5.6), a pyrome-
ter must be adjusted for the transmission properties of any supplemen-
tary lens and/or protective window screwed onto the pyrometer. Set the
pyrometer for the specific transmission factor (a percentage value) which
is indicated either in the products's data sheet or on the lens itself. This
TAU.1
is parameter
and can be accessed at configuration layer
C002/C003
. If you are not using any auxiliary lens or protective win-
dow, set the parameter to 100.0.
Ambient Temperature Compensation (Spectral mode)
Under normal conditions, PA pyrometer should produce highly accurate
temperature readings once you have set the emissivity and transmit-
tance factors. Certain conditions, however, will require that you make an
additional correction. Object surfaces which have very low emissivity will
reflect thermal radiation from their surroundings. If your target object is in
surroundings which are hotter than the target itself, and the infrared radi-
ation coming from extraneous sources reflects off of the object, the sen-
sor will collect this additional energy. To ensure that the measurement is
not distorted and that temperature readings are completely accurate, you
should activate „Ambient temperature compensation" (switch on
C002/C003 / BAC.1
)
Now enter the temperature value of the ambient source of radiation and
its influence as a percentage. You will have to determine these variables
first.
21