It is also possible to configure analogue output 2 with a sub-range which
covers a portion of the temperature span assigned to analogue output 1:
Configuration example PA 40:
Ao1: temperature reading of Lambda 1
650 - 1700 °C ≡ 4 - 20 mA
Ao2: temperature reading of Lambda 1
1000 - 1500 °C ≡ 4 - 20 mA
Digital outputs
One of the following functions can be assigned to each of the digital out-
puts:
Deactivate (This is required when you want to use the digital out-
put as a digital input).
Status LED lights up to indicate that the temperature reading of
Lambda 1 lies within the pyrometer's available temperature range.
Limit switch with adjustable signal threshold:
-
Quotient
-
Quotient before Min/Max before memory
-
Lambda 1
-
Lambda 1 before Min/Max
-
Lambda 2
-
Lambda 2 before Min/Max
-
Dirt Alert
-
Signal intensity
-
Device temperature
Status Signal ATD function.
-
Triggered by ATD function Lambda 1 at the end of the
measuring time
-
Triggered by ATD function Lambda 2 at the end of the
measuring time
-
Triggered by ATD function ratio mode at the end of the
measuring time
-
ATD function Lambda 1 indicate the measuring time
-
ATD function Lambda 2 indicate the measuring time
-
ATD function ratio mode indicate the measuring time