Digital Temperature

Knowing Your Temperature Controller (Tank 1, Tank 2, and Tank 3)
 PV: This upper display indicates the
actual Process Value (or current
temperature) of the bath and cannot
be changed manually.
 SV: The lower display indicates the
Set Value or desired bath
temperature. This Set Value is
adjustable.
 ALM 1: If this light is on, it means
that the machine is currently heating.
It will shut off once the desired
temperature has been reached.
 OUT: If this light is on, it means that
the bath temperature is out of the
preset range of 50˚F - 160˚F. There
is a 5˚ hysteresis which means that
the heating mechanism (and the
OUT light) will go on/off based on a
5˚ variance. This serves as a buffer
to protect the life of the motor
contactor.
 AT, ALM 2: Omegasonics use only.
 Return Key