 In the ultrasonic bath, preheat 1 part Omega AquaClean with 64 parts clean water to
a temperature of 115˚F. A higher dilution may be required depending on the severity
of contamination.
 If it exists, excess water should be removed as quickly as possible.
 UnWet CPDL should be used to stabilize the components if moisture exists, prior to
cleaning.
 Vacuum excess soot with a HEPA-style vacuum prior to placing in bath.
 Pre-spray the outside of the components with a 1 part OmegaSmoke, 64 parts clean
water mixture. Let sit for 2 – 5 minutes depending on the degree of the exterior soot.
 Place components in basket, do not stack, and clean for 2 minutes.
 Remove the basket from bath and rinse with free flowing de-ionized water.
 Blow off excess water with compressed air.
 Place a bowl of 1 part Omega Deodorizer and 1 part clean water at the bottom of the
drying chamber and dry the electronics for 10 hours at 120˚F.
 Ensure that the items are dry before connecting to a power source.