Printing Problems - Fluke Datapaq Food Tracker User Manual

Table of Contents

Advertisement

Proceed as follows.
1. Note readings first at ambient temperature: thermocouples registering no
data in Insight, or an open circuit with a digital thermometer (*OC* in the
Communications Setup dialog), may be broken. Inconsistent readings may
indicate an intermittent short circuit.
2. If a satisfactory ambient reading is recorded, apply heat to the
thermocouple-tip via fingers or other heat source. An increased
temperature should register:
○ If the reading does not change, the thermocouple is short circuit and
must be replaced.
○ If the probe measures air temperature, the cable may have damage which
has created a new hot junction.
○ If the thermometer shows a decrease, the thermocouple connections are
reversed.
3. Confirm correct operation at 100°C/212°F by placing the thermocouple-tip
in freshly-boiled water.
4. Replace any damaged cables.

Printing Problems

• Check correct printer selected: on menu bar, select File > Print Setup.
• Check printer cable connections.
Datapaq Service Department
If you cannot resolve your problem, please contact the Service Department at
Datapaq (see title page for contact details).