Anaesthetic Instruments (An) - Miele PG 8592 Operating Instructions Manual

Table of Contents

Advertisement

Areas of application
44

Anaesthetic instruments (AN)

The
Vario TD AN
programme with thermal disinfection is intended for
the reprocessing of anaesthetic instruments.
 Risk of heat damage.
The permissible reprocessing temperature is below 85 °C for some
elastomers used in breathing bags and breathing masks.
Observe the manufacturer's information on the permissible
reprocessing temperature to prevent the material from aging
prematurely.
Anaesthetic instruments should only be reprocessed in mobile
injector units designed for them.
Mobile units are supplied with their own operating instructions.
 If the process is not to be followed by sterilisation, the load
should be dried completely before storage to avoid the
development of water-borne bacteria.
The drying result must therefore be checked at the end of a
programme.
Tube interiors in particular must be completely dry. The programme
drying time must be adjusted as necessary to achieve this.