Application technology
Probes,
tweezers etc.
40
Instruments
Any deposits such as dental cement, composite, polishing paste or
similar must be removed immediately after examination of the patient,
e.g with a swab, before it hardens.
Instruments with particularly complex functional ends or very
stubborn deposits may require ultrasonic pre-treatment.
To avoid injury from double-ended instruments or upright
instruments with upward-facing probes, the washer-disinfector
should be loaded from rear to front, and unloaded from front to
rear.
After thermal disinfection, manual secondary cleaning can be done
according to relevant bio-substance regulations without danger of
infection, although the process must be monitored as necessary.
Mouth specula
Not all glass mouth specula can be reprocessed by machine.
Always follow the manufacturer's instructions.
Rhodium-coated mouth specula, because of their delicate surface,
must be loaded in such a way that the mirror surfaces cannot sustain
mechanical damage during reprocessing, e.g. by bumping against
other instruments.