3.3 Basic Waveform Output

3.3.1 Frequency Setting

Default waveform: A sine wave of 1kHz frequency and 100mV amplitude (with 50Ω termination).
Steps for changing the frequency to 2.5MHz are shown as following:
a)
Press Menu→Waveform→Parameter→Frequency in turn to frequency setting mode. Set parameters by
pressing Frequencysoftkey to change frequency and period.
b)
Use number keyboard to input the required number of 2.5.
c)
Select corresponding unit MHz.

3.3.2 Amplitude Setting

Defaultwaveform: A sine wave of 100mV peak-peak value with 50Ω termination.
Steps for changing the amplitude to 300mV are shown as following:
1.
Press Menu→Waveform→Parameter→Amplitude in turn. Press Amplitudesoftkey again can switch between
Vpp, Vrms, and dBm.
2.
Use number keys to input 300.