Setting The Function, Range, And Resolution - Agilent Technologies 34901A User Manual

Data acquisition/switch unit
Table of Contents

Advertisement

Chapter 5 Remote Interface Reference

Setting the Function, Range, and Resolution

CONFigure? [(
Query the present configuration on the specified channels and return a
series of quoted strings. If you omit the optional ch_list parameter,
the instrument uses the active scan list. The command returns a series
of comma-separated fields as shown in the example below. Each field
contains the function, range, and resolution.
"FRES +1.000000E+02,+3.000000E-04" , "TEMP TC,K,+1.000000E+00,+3.000000E-06"
Setting the Function, Range, and Resolution
See also "General Measurement Configuration" in chapter 4 starting
on page 98.
• For 4-wire measurements, the instrument automatically pairs
channel n with channel n+10 (34901A) or n+8 (34902A) to provide the
source and sense connections. Specify the paired channel in the lower
bank (source) as the ch_list channel.
• Current measurements are allowed only on channels 21 and 22
on the 34901A multiplexer module.
[SENSe:]FUNCtion "<
Select the measurement function on the specified channels. The function
name must be enclosed in quotes in the command string (for example,
FUNC "VOLT:DC"). Specify one of the following strings to set the function.
TEMPerature
VOLTage[:DC]
VOLTage:AC
RESistance
FRESistance
• Note that when you change the measurement function on a channel,
all of the other measurement attributes (range, resolution, etc.) are
set to their default values.
• You cannot set any function-specific measurement attributes unless
the channel is already configured for that function. For example,
you cannot set the ac filter unless that channel is already configured
for ac voltage or ac current measurements.
214
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
@
<
ch_list
>)]
function
@
>"[,(
<
CURRent[:DC]
CURRent:AC
FREQuency
PERiod
ch_list
>)]

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents