Agilent Technologies 34901A User Manual page 397

Data acquisition/switch unit
Table of Contents

Advertisement

Chapter 8 Tutorial
Multifunction Module
Using an External Pull-Up
In general, an external pull-up is required only when you want to set
the output "high" value greater than TTL levels. For example, to use a
+12V external power supply, the value of the external pull-up resistor is
calculated as follows:
V
= 12 Vdc
cc
I
= I
max
Vcc
R =
I
max
The value of the logic "high" level with the external 24 k
resistor is calculated as follows:
V
= V
high
Driving External Switches
You can use two digital output channels to control an external switch.
For example, you can drive the Agilent 876X series of microwave
switches using an external power supply and two digital output
channels. The state of the 2-to-1 multiplexer is changed by setting
the appropriate output bit low (0).
Digital Output Channel
394
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
x safety factor = 1 mA x 0.5 = 0.5 mA
out low
12
=
= 24 kΩ
0.0005
R
external
x
= 12 x
cc
+ R
R
external
internal
Agilent 876X Microwave Switch
24 Volt
Power
Supply
pull-up
24 k
= 8.47 Vdc
+
10
24 k
k
+