Scanning
Automatic Channel Delays
If you do not specify a channel delay, the instrument selects a delay
for you. The delay is determined by function, range, integration time,
and ac filter setting as shown below.
DC Voltage, Thermocouple, DC Current (for all ranges):
Integration Time
PLC > 1
PLC
1
Resistance, RTD, Thermistor (2- and 4-wire):
Channel Delay
Range
(For PLC > 1)
100
2.0 ms
1 k
2.0 ms
10 k
2.0 ms
100 k
25 ms
1 M
30 ms
10 M
200 ms
100 M
200 ms
AC Voltage, AC Current (for all ranges):
AC Filter
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Frequency, Period:
AC Filter
Slow (3 Hz)
Medium (20 Hz)
Fast (200 Hz)
Digital Input, Totalize:
Channel Delay
0 sec
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Channel Delay
2.0 ms
1.0 ms
Channel Delay
Range
(For PLC
100
1 k
10 k
100 k
1 M
10 M
100 M
Channel Delay
7.0 sec
1.0 sec
120 ms
Channel Delay
0.6 sec
0.3 sec
0.1 sec
1)
1.0 ms
1.0 ms
1.0 ms
20 ms
25 ms
200 ms
200 ms