Agilent Technologies 34901A User Manual page 288

Data acquisition/switch unit
Table of Contents

Advertisement

Chapter 5 Remote Interface Reference
The SCPI Status System
The Standard Operation Register
The Standard Operation register group is used to report when the
instrument is scanning. Any or all of these conditions can be reported to
the Standard Operation summary bit through the enable register. To set
the enable register mask, you must write a decimal value to the register
using the STATus:OPER:ENABle command.
Bit Definitions – Standard Operation Register
Bit Number
0
Not Used
1
Not Used
2
Not Used
3
Not Used
4
Scan in Progress
5
Not Used
6
Not Used
7
Not Used
8
Configuration Change
9
Memory Threshold
10 Not Used
15 Not Used
The Standard Operation event register is cleared when:
• You execute the *CLS (clear status) command.
• You query the event register using the STATus:OPER:EVENt? command.
The Standard Operation enable register is cleared when:
• You turn on the power (*PSC does not apply).
• You execute the STATus:PRESet command.
• You execute the STATus:OPER:ENABle 0 command.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Decimal
Value
Definition
1
Returns "0".
2
Returns "0".
4
Returns "0".
8
Returns "0".
16
Instrument is scanning (
32
Returns "0".
64
Returns "0".
128
Returns "0".
256
Channel configuration was changed from the
front panel. This bit is cleared when a new
scan is initiated.
512
Programmed number of readings have been
stored in reading memory.
1024
Returns "0".
32768
Returns "0".
SCAN
annunciator is on).
285
5

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents