State Storage Commands
MEMory:STATe
:NAME {1|2|3|4|5} [,<
:NAME? {1|2|3|4|5}
Assign a name to the specified storage location (you cannot assign a
name to location "0"). You can name a location from the front panel or
over the remote interface but you can only recall a named state from the
front panel. From the remote interface, you can only recall a stored state
using a number (0 through 5). The :NAME? query returns a quoted
string containing the name currently assigned to the specified storage
location. If the specified location has no name assigned, an empty string
(" ") is returned.
• The name can contain up to 12 characters. The first character must
be a letter (A-Z), but the remaining 11 characters can be letters,
numbers (0-9), or the underscore character (" _ "). Blank spaces are
not allowed. An error is generated if you specify a name with more
than 12 characters. An example is shown below.
MEM:STATE:NAME 1,TEST_RACK_1
• If you do not specify a name (note that the name parameter is
optional), no name is assigned to that state. This provides a way to
erase a name (however, the stored state is not deleted).
MEMory:STATe:DELete {0|1|2|3|4|5}
Delete the contents of the specified storage location. If you have named
a storage location (MEM:STAT:NAME command), this command does not
remove the name that you assigned. Note that you cannot recall the
instrument state from a storage location that was deleted. An error is
generated if you attempt to recall a deleted state.