Chapter 4 Features and Functions
General Measurement Configuration
• Remote Interface Operation: Specify the resolution in the same units
as the measurement function, not in number of digits. For example,
if the function is dc volts, specify the resolution in volts. For frequency,
specify the resolution in hertz.
You can select the resolution using parameters in the MEASure? and
CONFigure commands. For example, the following statement selects
the 10 Vdc range with 4
CONF:VOLT:DC 10,0.001,(
The following statement selects the 1 A range with 6
resolution on channel 221.
MEAS:CURR:AC? 1,1E-6,(
You can also select the resolution using the SENSe commands.
For example, the following statement specifies a 4-wire ohms
measurement with 100
SENS:FRES:RES 100,(
102
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
digits of resolution on channel 301.
2
@
301)
@
221)
of resolution on channel 103.
@
103)
1
digits of
2