Agilent Technologies 34901A User Manual page 103

Data acquisition/switch unit
Table of Contents

Advertisement

Chapter 4 Features and Functions
General Measurement Configuration
Measurement Resolution
Resolution is expressed in terms of number of digits the instrument can
measure or display on the front panel. You can set the resolution to
4, 5, or 6 full digits, plus a "
To increase your measurement accuracy and improve noise rejection,
select 6
1
5 digits
1 0 . 2 1 6 , 5
1
"
" digit
2
This is the 10 Vdc range, 5
1
"
" digit
2
-045.23
This is the 100 mVdc range, 4
113.325,6 OHM
This is the 100 ohm range, 6
• For temperature measurements taken from the remote interface,
the resolution is fixed at 6
the resolution in terms of the number of digits displayed past the
decimal point (Measure menu).
• For ac voltage measurements, the resolution is fixed at 6
The only way to control the reading rate for ac measurements is by
changing the channel delay (see page 88) or by setting the ac filter to
the highest frequency limit (see page 114).
100
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
1
" digit which can only be a "0" or "1".
2
digits. To increase your measurement speed, select 4
2
V D C
1
digits are displayed.
2
mVDC
1
digits are displayed.
2
1
digits are displayed.
2
1
digits. From the front panel, you can set
2
1
digits.
2
1
digits.
2