Instructions for Use
Remove O-rings carefully with your fingers. Do not use metal tools for this
CAUTION
task; this prevents scratching of any sealing surfaces.
• Wipe all O-rings clean with a lint-free cloth before installation to ensure that no
foreign matter is present to impair the seal.
• Do not use grease or any other substance on O-rings that will come in contact
with the spectrometer.
• Do not use alcohol, methanol or other solvents on O-rings. Doing so causes
deterioration and reduces their ability to hold a vacuum.
• Agilent does not recommend the use of vacuum grease. If applicable, apply a
small amount of Apiezon® L grease and wipe the O-rings shiny dry.

Metal Seal Care

Metal Seals must be replaced during routine maintenance or during any
CAUTION
maintenance procedure required. All fasteners must be installed and torqued per
assembly procedure specifications. Remove Metal Seals carefully with your
fingers. Do not use metal tools for this task, this prevents scratching of any
sealing surfaces. Wear non-powdered, ESD safe Nitride or equivalent gloves prior
to removing or replacing Metal Seals.
• All Metal Seals come in clean room condition packages. No cleaning is
required. If necessary, Metal Seals can be cleaned using alcohol or methanol.
Wipe all Metal Seals clean with a lint free cloth before installation to ensure
that no foreign matter is present to impair the seal.
• Do not use grease or any other substance on Metal Seals that will come in
contact with the spectrometer.