7.9.4 Gravitational constant adjustment location < grause >
22
Press and hold the TARE and ON/OFF buttons
simultaneously to enter the setup menu.
Wait until the first menu item <Cal > is displayed.
Confirm by ➔ button, <CalExt > will be displayed.
Use the navigation keys to select   < grause >.
Acknowledge using ➔ button, the current setting is
displayed. The active digit is flashing.-+
Enter weight value and confirm using the ➔ button,
numeric entry see chap. 3.2.2.
Weighing balance returns to menu.
Press repeatedly  button to exit menu.