Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual

General

Calibration Number and Date

When storing valid calibration data in FlashROM after the calibration adjustment
procedure is complete, the calibration date is set to the actual Test Tool date,
and the calibration number increments by one. To show the calibration date and
number:
1. Press
(Figure 5-1).
2. Press
The calibration date and calibration number do not change if you do
only the Contrast Calibration Adjustment and/or the Probe
Calibration.

General Instructions

Follow these general instructions for all calibration steps:
Allow the specified warm-up period for the 5502A. For each calibration point,
wait for the 5502A to settle.
The required warm-up period for the Test Tool is included in the Warming-Up
and Pre-Calibration step.
Ensure that the Test Tool battery is charged sufficiently.
Power the Test Tool with the BC190 Power Adapter.
This procedure is for all models. Test steps that are not applicable to the Test
Tool to be adjusted can be skipped. For example, the adjustment of the meter
with banana jacks can be skipped in instruments with four scope (BNC)
inputs.
The figures that show how to interconnect Signal Source and Test Tool are
for 2 Scope Inputs + Meter Input and for 4 Scope Inputs.
5-4
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
USER
F3
, then press
F4
to close the version and calibration menu.
Figure 5-1. Version and Calibration Data (example)
Note
to see the version and calibration data
version-calibration-3.bmp

Advertisement

Table of Contents
loading

Table of Contents