Introduction - Fluke ScopeMeter 190-202 Service Manual

Hide thumbs Also See for ScopeMeter 190-202:
Table of Contents

Advertisement

Introduction

To prevent possible electrical shock, fire, or personal injury, do
not service the Test Tool unless you are qualified to do so.
Service described in this manual is to be done only by qualified
service personnel.
Table 4-1 lists the available models for the Test Tools.
Model
190-062
Two 60 MHz Scope Inputs (BNC), one Meter Input (banana jacks).
190-102
Two 100 MHz Scope Inputs (BNC), one Meter Input (banana jacks).
190-104
Four 100 MHz Scope Inputs (BNC).
190-202
Two 200 MHz Scope Inputs (BNC), one Meter Input (banana jacks).
190-204
Four 200 MHz Scope Inputs (BNC).
190-502
Two 500 MHz Scope Inputs (BNC), one Meter Input (banana jacks).
190-504
Four 500 MHz Scope Inputs (BNC).
The Test Tool should be calibrated and in operating condition on arrival.
The following performance tests are provided to ensure that the Test Tool is in
proper operating condition. If the Test Tool fails any of the performance tests,
calibration adjustment (see Chapter 5) and/or repair is necessary.
The Performance Verification Procedure is based on the specification (see
Chapter 2). The values given here are valid for ambient temperatures between
18 °C and 28 °C.
The Performance Verification Procedure is a quick and efficient way to check all
main specifications of the Test Tool. Accuracy of Test Tool specifications not
tested is linked to those tested in this verification procedure and is embedded in
the Test Tool's software. This link is tested extensively for each new software
release.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
 Warning
Table 4-1. Model Descriptions
Description of Main Features
Performance Verification

Introduction

4
4-3