Input B LF-HF Gain

To do the Input B LF-HF Gain calibration:
F2
1. Press
2. Connect the Test Tool to the 5502A as shown in Figure 5-12.
PM9091
50 OHM e.g. PM9585 or
TRM50
For 190-502 and -504
use only TRM50!
METER
A
B
A
2 CHANN. + METER
Figure 5-12. 5502A SCOPE Output to Test Tool Input B
3. Set the 5502A SCOPE output to source the signal required for the first
calibration point in Table 5-11.
4. Set the 5502A to operate (OPR) or standby (STBY) as indicated.
F3
5. Press
6. Wait until the display shows calibration status
F2
7. Press
calibration point signal, and start the calibration.
Continue through all calibration points of Table 5-11 and Table 5-12.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Input C LF-HF Gain section.