Fluke ScopeMeter 190-202 Service Manual page 112

Hide thumbs Also See for ScopeMeter 190-202:
Table of Contents

Advertisement

F3
5. Press
to start the calibration.
6. Wait until the display shows the calibration status
F2
7. Press
to select the next calibration step, set the 5502A to the next
calibration point, and start the calibration. Continue through all calibration
points.
8. When you are finished, set the 5502A to Standby.
9. Continue at the Save Calibration Data and Exit section.
Table 5-8. Ohm Gain Calibration Points
Cal Step
CL 0910
CL 0911
CL 0912
CL 0913
CL 0914
CL 0915
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Calibration Adjustment
Final Calibration For v11.10 and later
.
:READY
UUT input Value (5500 NORMAL)
100 Ω
1 kΩ
10 kΩ
100 kΩ
1 MΩ
10 MΩ
5
5-27