Fluke ScopeMeter 190-202 Service Manual page 117

Hide thumbs Also See for ScopeMeter 190-202:
Table of Contents

Advertisement

Fluke 190-062, -102, -104, -202, -204, -502, -504
Service Manual
Cal step
CL 0674
CL 0482
CL 0483
Table 5-12. Input B LF-HF Gain Calibration Points (Extra steps for 190-104/204, V10.4x firmware *)
Cal step
CL 0462
CL 0463
* Four channel instruments (190-104/204) with firmware V10.4x require the extra steps CL 462 and CL 463. Installed
firmware version can be checked via key sequence: USER key, F3 – VERSION & CAL, Software Version: ...
5-32
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
Table 5-11. Input B LF-HF Gain Calibration Points
UUT input signal
0.5 Vpp square wave, 1 kHz
0.62 Vpp sine wave, 50 kHz
0.62 Vpp sine wave
Fluke 190-202/204:
221 MHz
Fluke 190-102:
151 MHz
Fluke 190-104:
121 MHz
Fluke 190-062:
81 MHz
UUT input signal
124 mVpp sine wave, 50 kHz
124 mVpp sine wave
Fluke 190-204:
221 MHz
Fluke 190-104:
121 MHz
5502A Setting
SCOPE edge, 0.5 Vpp, 1 kHz
SCOPE levsine, 0.62 Vpp, 50 kHz
SCOPE levsine, 0.62 Vpp,
221 MHz
151 MHz
121 MHz
81 MHz
5502A Setting
SCOPE levsine, 124 mVpp, 50 kHz
SCOPE levsine, 124 mVpp,
221 MHz
121 MHz

Advertisement

Table of Contents
loading

Table of Contents