zero ref- erence.
Wait for the temperature reading displayed on the Temperature Calibration dialog box to be sta-
bilized.
3. After the reading stabilizes, click Average Temperature Now .
MasterLinkR515 &  MasterLinkSQL averages the reading and displays the average temperature.
4. Click Change.
The Change Temperature Calibration Value dialog box appears.
5. Change the value to accurately match the temperature measured by the calibrated thermometer
placed in the low temperature bath.
6. Click OK.
7. Click Span Calib.
The Temperature Span Calibration page appears.
8. Insert the temperature probe of ERX350 and the calibrated thermometer into a high temperature
bath, which provides the temperature span reference.
The temperature reading displayed on the Temperature Calibration dialog box must exceed the zero
reference temperature by at least 15% to proceed with span calibration.
9. After the reading stabilizes, click Average Temperature Now.
MasterLinkR515 &  MasterLinkSQL averages the reading and displays the average temperature.
Honeywell
2023
5.4   Working with MasterLinkR515 &  MasterLinkSQL