SCPI Command Reference

ABORt

Subsystem Syntax
Comments
Example
44 HP E1364A Form C Switch Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
This section describes the Standard Commands for Programmable
Instruments (SCPI) commands for the Form C Switch. Commands are listed
alphabetically by subsystem and also within each subsystem.
The ABORt command subsystem stops a scan in progress when the scan is
enabled via the interface and the trigger source is TRIGger:SOURce BUS
or TRIGger:SOURce HOLD.
ABORt
ABORt Actions: ABORt invalidates the current channel list and sets
ARM:COUNt 1 (one scanning cycle per INITiate command), sets
INITiate:CONTinuous OFF (no continuous scanning cycles), and
sets TRIGger:SOURce IMMediate (immediate internal triggering).
Stopping Scan Enabled From Interface: When a scan is enabled
from an interface, an interface CLEAR command or the HP E130lA
front panel "Reset Instr" or "Clear Instr" keys can be used to stop the
scan. When the scan is enabled from the interface and TRIG:SOUR
BUS or HOLD is set, use ABORt or the HP E130lA front panel
"Reset Instr" or "Clear Instr" keys to stop the scan.
Stopping Scan Enabled From Front Panel: When a scan is
enabled from the HP E1301A front panel, you can use *RST entered
via the interface or the front panel "Reset Instr" or "Clear Instr" keys
to stop the scan.
Related Commands: ARM, INITiate:CONTinuous,
[ROUTe:]SCAN, TRIGger
Stopping a scan with ABORt:
This example stops a (continuous) scan in progress.
TRIG:SOUR BUS
INIT:CONT ON
SCAN (@100:103)
INIT
.
.
ABOR
*TRG command is trigger source
Set continuous scanning
Scan channels 00 through 03
Start scan, close channel 00
Abort scan in progress
Chapter 5