:CPON
Parameters
Comments
Example
TRlGger
Subsystem Syntax
[:IMMediate]
Comments
58 HP E1364A Form C Switch Command Reference
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com
SYSTem:CPON < number > | ALL
switchbox to its power-on state.
Parameter Name
Parameter Type
< number >
numeric
•
Form C Switch Module Power-On State: The power-on state is all
channels (relays) open. Note that *RST opens all channels of all
modules in a switchbox, while SYSTem:CPON <number> opens the
channels in only the module (card) specified in the command.
Setting Card #1 Module to its Power-On State:
SYST:CPON 1
The TRIGger command subsystem controls the triggering operation of
Form C Switch modules in a switchbox.
TRIGger
[:IMMediate]
:SOURce BUS | EXTernal | HOLD | IMMediate
:SOURce?
causes a trigger event to occur when the defined
TRIGger[:IMMediate]
trigger source is TRIGger:SOURce BUS or TRIGger:SOURce HOLD.
•
Executing the TRIGger[:IMMediate] Command: A channel list
must be defined with [ROUTe:]SCAN <channel_list> and an
INITiate[:IMMediate] command must be executed before
TRIGger[:IMMediate] will execute.
•
BUS or HOLD Source Remains: If selected, the TRIGger:SOURce
BUS or TRIGger:SOURce HOLD commands remain in effect after
triggering a switchbox with the TRIGger[:IMMediate] command.
•
Related Commands: INITiate, [ROUTe:]SCAN
sets the selected module (card) in a
Range of Values
1 - 99
Sets module #1 to power-on state
Chapter 5