Thank you for purchasing the Welch Allyn Video Colposcope. Follow the operation and maintenance instructions found in this manual and your video colposcope will provide you with many years of reliable service. Please read these instructions thoroughly before attempting to use your new video colposcope.
The lamp is extremely bright. DO NOT stare directly into illumination lens when the lamp is lit. • All service to the video colposcope must be performed by Welch Allyn or by an authorized repair center. • Video colposcope user should adhere to the operating conditions found in this manual. Otherwise, instrument damage may occur and/or operator/patient safety may be compromised.
The CE mark on this product indicates it has been tested to and conforms with the provisions noted in the 93/ 42/EEC Medical Device Directive. Authorized European Representative: European Regulatory Manager Welch Allyn, Ltd. Kells Road, Navan, County Meath, Republic of Ireland Tel 353 46 79060...
Maintenance section of this manual. WARNING: The lamp operates at a high temperature. DO NOT attempt to remove the lamp before allowing it to cool. Allow at least five minutes for the lamp to cool before replacing. Replace with Welch Allyn lamp #09800 only.
Components Video Colposcope 88000A 89000A Video colposcope and vertical stand with Video colposcope and swing arm stand vertical stand hardware kit: with swing arm hardware kit: • Allen wrench • Allen wrench • Allen bolt • Bolt • Mounting washer •...
Page 10
Dust Cover for Vertical Stand 89040 Dust Cover for Swing Arm 88030 Vertical Stand with Base Only 89030 Swing Arm Stand with Base Only NOTE: Only accessories and components indicated in this manual are to be used with the Welch Allyn Video Colposcope system.
Nomenclature Video Colposcope Front View Name Function Objective Lens The lens that establishes the magnification and field of view. Illumination beam director Directs lamp beam. Video Colposcope Side Views Name Function Ventilation slots Allow air to enter and leave unit, cooling internal components. Zoom control button Pressing + or - increases or decreases magnification accordingly.
Video Colposcope Back View Name Function 11. Power switch Power control for the video colposcope. 12. Ventilation slots Allow air to enter and leave unit, cooling internal components. 13. Video colposcope handle For positioning the colposcope in the proper up/down, left/right position. 14.
Vertical Pole Stand Name Function 19. Vertical rolling base stand Allows mobility with 5-caster base, including two locking casters. 20. Fuse drawer Opens for replacement of fuses. 21. Power supply cord receptacle Couples with power cord, which should be plugged into a 110-120 VAC hospital-grade outlet (220-240 volt, 50 cycle international).
Monitor NOTE: The following is a brief overview of the nomenclature of the Sony PVM-14N1U Video Monitor. For detailed instructions, please refer to the operation manual that has been provided by the manufacturer Front Panel of Monitor Back Panel of Monitor PLEASE REFERENCE YOUR SONY MONITOR MANUAL FOR FURTHER INFORMATION.
WARNING: All video monitors and peripheral equipment used with this instrument must be IEC 601-1 (or UL 2601) listed to comply with medical electrical safety guidelines. If not IEC 601-1(UL-2601) listed, the monitor/ peripheral equipment should be positioned outside of the 6-foot patient contact area. Please contact Welch Allyn for more information.
Before initial set up of the video colposcope, check all components received against the parts list of components (see Compo- nents section of this manual) to verify a complete set. If parts are missing, please notify Welch Allyn. Review the Nomencla- ture, Preparation for Use, Operation, and Maintenance sections to become familiar with the equipment.
Place pole into the base, aligning the pin on the flange with the pin hole in the base. Place bolt, spacer, and washer, ori- ented as shown, into the pole. Tighten securely with the enclosed Allen wrench. (The assembly may need to be tilted.) Washer Spacer Hex Bolt...
11. Bring the colposcope swing arm assembly to the pole while at the same time pushing the electrical connectors into the colposcope arm. 12. Tighten the bolt securely with the enclosed Allen wrench. Be sure not to pinch any of the electrical wires between the vertical pole and arm of the colposcope assembly.
Operation Power Switch With the power cord connected to a properly grounded outlet, and the S-video cable connected to the video colposcope and monitor, activate the power switch on the handle of the video colposcope. Lamp Ignition Once the video colposcope's power switch has been activated, the lamp will ignite. The lamp requires approximately 11 sec- onds to warm up.
Illumination Beam Director The illumination beam director can be rotated up to 45° clockwise or counterclockwise to better illuminate the examination area. To rotate the director, grasp the flat sides and twist in either direction as desired. The illumination beam director moves with some resistance.
References to brand names are not endorsements of their efficacy as disinfecting solutions. However, tests have shown these solutions to be compatible with Welch Allyn colposcopes, providing the manufacturers' directions are followed. Do not use any other solutions unless a sample has been sent to Welch Allyn for compatibility test- ing.
Replacing Lamp WARNING: The lamp operates at a high temperature. DO NOT attempt to remove the lamp before allowing it to cool. Allow at least five minutes for the lamp to cool before replacing. Replace with Welch Allyn lamp #09800 only.
Page 23
Remove a Welch Allyn replacement lamp # 09800 from its package. Do not touch the lamp itself or the interior reflective surface of the lamp. Skin oils will cause premature lamp failure. Hold lamp by the outside of the reflector or the connector only.
Pull out and remove the blown fuse from the fuse holder. Replace with new fuses # T1.00L-250V 1A Time lag/Low Breaking Capacity (Welch Allyn Part # 488307-9). Reinsert the fuse holder by pressing until it snaps into place.
U.S.A. Telephone: 1-800-535-6663 Fax: 315-685-4653 Customers outside of North America should return their units to a local, authorized Welch Allyn distributor, or to their nearest Welch Allyn service center. Welch Allyn, Inc., U.S.A. - 4341 State Street Road, Skaneateles Falls, NY 13153-0220, Tel: 800-535-6663, Fax: 315-685-4653 Welch Allyn, Ltd., Canada -...
Make sure lamp access door and handle are prop- and door to unit erly attached. Fuses Remove fuse panel & replace blown fuse with T1.00L 250V 1A time lag (Welch Allyn part number # 488307-9). Wall outlet Plug the power cord into a wall outlet known to work.
21 Watt metal halide arc Lamp life 750 hrs @ 1 hr per start Brightness adjustment Automatic electronic shutter Lamp Welch Allyn part no. 09800 Lamp voltage 60 volts Cooling method Forced air via fan Color system Color mosaic CCD...
Page 28
IEC601-1-2 Conformace Explosion proofing Use with potentially flammable surroundings is prohibited Fuses T1.00L-250V 1 Amp time lag, (Welch Allyn part # 488307-9) low breaking capacity Classification Electro medical equipment Class I equipment Degree of protection against Type B electrical shock...