Advertisement

5.
Push the retainer spring toward the back of the
colposcope and remove the lamp from its housing.
6.
Remove a Welch Allyn replacement lamp # 09800 from its package. Do not touch the lamp itself or the interior reflective
surface of the lamp. Skin oils will cause premature lamp failure. Hold lamp by the outside of the reflector or the connector
only. Remove any grease or fingerprints with a clean cotton swab moistened with alcohol. Do not leave any lint on the
lamp.
7.
Holding the retainer spring as in 'step 5' , slide the new lamp into the lamp housing so that the lamp's alignment pin
engages the alignment slot in the lamp holder. (Make sure lamp is properly seated and snaps into place.)
8.
Plug the electrical connector onto the new lamp.
9.
Slide the lamp access door securely back into place, reversing the process described in 'step 3' above.
10. Place the video colposcope back into the handle, mak-
ing sure the power connector is engaged.
11. Tighten the knob by rotating clockwise until snug.
12. Plug the power cord back into the electrical outlet.
19