Table 33 – Tamper Reporting Logic Implementation with Write Protect
ON
OFF (or missing)
OFF (or missing)
ON
OFF
Note that Tamper Reporting is independent of Write Protect status.
The sections below give some examples as to how to edit the configuration parameters and execute
Methods.
NOTE:
The following sections detail some of the basic operations of FDC applications. After writing
dynamic parameters like Ranges, units, limits and conversion factors (as applicable) to the device,
close the device and load the device again to see the new values. Sometimes screen refresh may take
some time before updating the new values. Reloading should refresh all the screens.

Procedure to Enter the Transmitter Tag

1. From the My Device menu, make the following menu selections:
Device Setup > Basic Setup > Device Information > Tag.
2. Click Edit. The Tag screen will be displayed.
3. Key in the tag name (for example: SMV800) which can be a maximum of eight characters.
4. Click OK. The Send to Device screen will be displayed.
5. Select the Tag check box.
6. Click Send to download the change to the Transmitter, or Click Return to continue making
changes.

Selecting Variable units of measurement

Process Variable (PV), Secondary Variable (SV), Tertiary Variable (TV), Quaternary Variable
units of measurement
See –
Revision 6.0
Write Protect
Jumper Status
Software Status
ON or OFF
ON
OFF
Tamper Reporting
Status
SMV800 Series HART/DE Option User's Manual
Configuration
Write Protect
Allowed?
NO
NO
YES
Tamper Alerted Posted?
YES
NO
Change
Page 121