Prerequisites; Objective; Required Equipment; Procedure - Siemens MAMMOMAT Inspiration Quality Control Manual

Table of Contents

Advertisement

Prerequisites

Objective

To determine whether the detector is damaged or has dust on it and is calibrated.

Required Equipment

a)
Lint-free, non-woven cotton or gauze (100% cotton)
b)
Water or lukewarm diluted aqueous solution of household dishwashing liquid

Procedure

1.
Wipe the detector surface with a wet lint-free, non-woven cotton cloth or cotton
(100%) pad. For moistening, use water or lukewarm diluted aqeous solution of
household dishwashing liquid.
NOTE!
NOTE!
2.
Calibrate the detector according to the chapter Calibrating the detector in the
MAMMOMAT Inspiration Operator Manual.
CAUTION!
CAUTION!
CAUTION!

Performance Criteria and Corrective Action

No visual damage to the detector should be observable.
No artifacts should be visible.
MAMMOMAT Inspiration
SPB7-330.640.01.01.02
Do not spray the unit! The cleaning fluid must under no circumstances
penetrate into the unit.
Improper operation can lead to reduced image quality.
Let the MAMMOMAT Inspiration warm up for at least 30 minutes to
ensure optimum results when performing an image quality test or a
calibration.
For startup and login, see MAMMOMAT Inspiration Operator Manual.
Before starting the quality control tests the detector calibration shall be
performed to ensure that the tests are based on correct detector
settings, see MAMMOMAT Inspiration Operator Manual.