MAGSTIM 3190-00 Operating Manual page 34

2nd generation coils
Hide thumbs Also See for 3190-00:
Table of Contents

Advertisement

Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Magstim 200
or the user of the equipment should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-4
Electrical fast
Transient/ burst
IEC 61000-4-11
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions and
voltage variations
on power supply
input lines.
IEC 61000-4-11
Power Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
Immunity
EN61000-4-3
Conducted RF
Immunity
EN61000-4-6
U
NOTE
is the a.c. mains voltage prior to application of the test level
T
© The Magstim Company Limited
2
is intended for use in the electromagnetic environment specified below. The customer
IEC 60601
Test Level
± 6 kV contact
Meets requirement
± 8 kV air
± 2 kV for power
Meets requirement
supply lines
± 1 kV for input/
output lines
± 1 kV differential
Meets requirement
mode
± 2 kV common
mode
U
<5 %
Meets requirement
T
U
(>95 % dip in
)
T
for 0,5 cycle
U
40 %
T
U
(60 % dip in
)
T
for 5 cycles
U
70 %
T
U
(30 % dip in
)
T
for 25 cycles
U
<5 %
T
U
(>95 % dip in
)
T
for 5 sec
3 A/m 50Hz
Meets requirement
80MHz – 2.5GHz
Meets requirement
2Hz 80%
amplitude
modulation
0.15MHz – 80MHz
Meets requirement
2Hz 80%
amplitude
modulation
Compliance
Electromagnetic environment
Level
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Magstim 200
continued operation during
power interruptions, it is
recommended that the Magstim
200
interruptible power supply.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
31
Magstim 2nd Generation Coils
Operating Manual
- guidance
2
requires
2
be powered from an
MOP04-EN-03