Guidance and Manufacturer's Declaration – Electromagnetic Immunity
The Magstim Rapid
The customer or the user of the equipment should assure that it is used in such an environment.
Immunity Test
Electrostatic
discharge (ESD)
IEC 61000-4-4
Electrical fast
Transient/ burst
IEC 61000-4-11
Surge
IEC 61000-4-5
Voltage dips,
short
interruptions
and voltage
variations on
power supply
input lines.
IEC 61000-4-11
Power
Frequency
(50/60Hz)
Magnetic Field
IEC 61000-4-8
Radiated RF
Immunity
EN61000-4-3
Conducted RF
Immunity
EN61000-4-6
NOTE
U
is the a.c. mains voltage prior to application of the test level
T
© The Magstim Company Limited
2
System is intended for use in the electromagnetic environment specified below.
IEC 60601
Test Level
± 6 kV contact
Meets requirement
± 8 kV air
± 2 kV for power
Meets requirement
supply lines
(Resetting of the MEP
Pod may be experienced
± 1 kV for input/
during single shot mode
output lines
under these conditions.)
± 1 kV differential
Differential mode
mode
compliant.
± 2 kV common
Common mode compliant
mode
to ± 1kV only.
<5 %
U
Meets requirement
T
(>95 % dip in
U
)
T
for 0,5 cycle
U
40 %
T
U
(60 % dip in
)
T
for 5 cycles
U
70 %
T
U
(30 % dip in
)
T
for 25 cycles
U
<5 %
T
(>95 % dip in
U
)
T
for 5 sec
3 A/m 50Hz
Meets requirement
80MHz – 2.5GHz
Coil adapter P/N 3110-00
2Hz 80%
and Cooled Coil P/N
amplitude
3530-00 between 80MHz
modulation
and 100MHz may
produce system errors
setting system into safe
mode.
0.15MHz – 80MHz
Coil adapter P/N 3110-00
2Hz 80%
and Cooled Coil P/N
amplitude
3530-00 between 30MHz
modulation
and 80MHz may produce
system errors setting
system into safe mode.
Compliance
Electromagnetic environment
Level
Floors should be wood, concrete
or ceramic tile. If floors are
covered with synthetic material,
the relative humidity should be at
least 30%.
Mains power quality should be
that of a typical commercial or
hospital environment.
Anti-surge protection needs to be
incorporated into the main supply
to the equipment if surge
protection is to be guaranteed.
Mains power quality should be
that of a typical commercial or
hospital environment. If the user
of the Magstim Rapid
requires continued operation
during power interruptions, it is
recommended that the Magstim
Rapid
an uninterruptible power supply.
Power frequency magnetic fields
should be at levels characteristic
of a typical location in a typical
commercial or hospital
environment
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
Equipment should only be used in
the vicinity of other equipment
compliant with EN60601-1-2.
27
Magstim 2nd Generation Coils
Operating Manual
- guidance
2