Storage and Shipment

CAUTION:
Do not store calorimeter or load where ambient temperatures fall below freezing.

Drain the calorimeter system of all coolant

1. Disconnect the inlet and outlet hoses from the chiller and 1314.
The hoses may be removed without spilling any fluid from either the 1314 or chiller as all
the hose connection ports have valves and prevent fluid from flowing when disconnected.
Drain the individual hoses completely by removing the connectors on their respective
ends.
2. Place a quick connect connector with an open hose end onto the return inlet of the chiller.
3.
Put the end of the open hose into a vessel to collect the fluid and loosen the reservoir cap.
4. Tilt the unit about 30 degrees to drain as much fluid as possible.
To completely empty the system put another connector and clean hose into the supply inlet
on the chiller. Apply low pressure air (less than 10 psi) for a minimum of 5 minutes or until
coolant is no longer expelled from the unit.
Repeat this procedure using the outlet hose port, and the input fluid port.
1314-900 Revision BA
250W Precision Calorimeter
30