Table 3: General Troubleshooting Guide
Symptom
No power displayed
Power won't settle or excessively long settling
time
Flow rate much less than nominal
AC calibrated power equals input power
RF calibrated power equals AC calibrated power
Extreme ambient temperature readings
No communication with 1314 Calorimeter
1314-900 Revision BA
Corrective Action
Check load DC resistance for possible failure
Check RF amp for possible power drifting
Make sure chiller set point and PID parameters
are correct
Check for blockage in fluid filter
Check calibration.ini file for proper values
Check reflection.ini file for proper values
Check for fan failure or inlet air blockage
Check for proper IP address
250W Precision Calorimeter