Installation and Set-Up
Unpack the Shipment and Inspect the Contents
Each calorimeter is put through a series of electrical and mechanical inspections before shipment
to the customer.
1. Upon receipt of your instrument unpack all the items from the shipping carton.
2. Inspect for damage that may have occurred during transit. Report any damaged items to the
shipping agent. Retain and use the original packing material for reshipment if necessary.
3. Inspect the carton for the following items:
Table 1: Packing List
Item
Model 1314 Precision Calorimeter
Technical (Operation and Maintenance) Manual
Chiller
Calibration Certificate with Data
Software
Calibration Cable
Optional Accessories
Control the Installation Environment
Normal calibration laboratory practice dictates the environment should be closely controlled. This
minimizes errors introduced by temperature and humidity changes. A nominal temperature of
+23ºC (+73.4ºF) provides a good working condition. A tolerance of ±1ºC gives an ideal
temperature spread. Controlled temperatures also stabilize the aging process of the standards.
1314-900 Revision BA
Part Number
1314-300
1314-840(CD) or 1314-900(Printed)
1314-389
n/a
HPC-CAL
1314-252
Various
250W Precision Calorimeter
3
8