Check the Location Where the Unit Will Be Installed

1. Make sure the surface is level. An unbalanced system rack can produce faults and/or
inaccuracies.
2. Select a spacious area with good airflow. Avoid placing the unit next to heat producing
equipment or in locations subject to drafts or sudden changes in temperature. Free flowing air
through the intake and outlet gratings on the rack are important. Notice the red and blue
arrows depicting air flow in Figure 4 below.
Figure 5: Airflow
3. Lock caster wheels once location is determined.

Fill the Coolant Reservoir

CAUTION:
Never operate cooling pump dry! Severe damage will occur.
Never apply AC or RF power to the unit until proper coolant levels and sufficient air
flow are established.
CAUTION:
Use only pure distilled water in the coolant mix.
Do not fill the unit with de-ionized water, flammable fluids, corrosive fluids, explosive,
or any other hazardous fluids.
1. Disconnect the main power cord!
1314-900 Revision BA
250W Precision Calorimeter