5. Once communication has been established, proceed to the parameter list shown in Appendix
A and check the following parameters to display their real time data being streamed from the
1314:
RTD_TEMP_PRE
RTD_TEMP_POST
THERMISTOR
FLOWRATE (GPM)
THERMOPILE
The RTDs and ambient values should now display real time data of values close to the current
laboratory environment (typically ~ 25C). The thermopile should display a value close to 0
(typically .0009 or so) as there is no temperature differential taking place yet with no fluid or flow
present. The flowrate should display 0 as there is no fluid present and the pump has not been
turned on yet. This flowrate value will guide you to establishing the proper flowrate once the
pump has been turned on. The unit is now ready to prime the pump and the proper flowrate will
be adjusted to a value of .5 to .55 GPM using the flow adjust knob on the rear of the 1314 unit as
described in the next section

Connect the Chiller and Prime the Coolant Pump

1. Connect the chiller serial cable to the PC USB port.
2. Ensure the supply outlet hose of the chiller is connected to a filter then the inlet of the 1314.
3.
Make sure the outlet hose of the 1314 is connected to the return inlet of the chiller.
4.
Locate the main power switch on the back side of the chiller, turn it ON.
1314-900 Revision BA
.
250W Precision Calorimeter
12