Advanced Energy 1314 User Manual page 9

Precision calorimeter 250w
Table of Contents

Advertisement

Storage Temperature
Warranty
S11 Linear Magnitude (Typical)
Optional System Components (contact your AE-TEGAM representative for more information).
AC Source
AC Power Meter
Rack (Single/Dual Bay)
P/C
Other
1
Average of ten (10) consecutive points taken 25 seconds apart, where the standard deviation of
those points is < 0.05 W.
2
The 1314 is intended for use as a primary standard. Measurement repeatability and uncertainties
will vary among laboratories depending upon various factors, including ambient environment,
user experience, and elapsed time since the most recent system calibration. TEGAM will provide
guidance where possible, but users are ultimately responsible for establishing their own
measurement uncertainties consistent with the laboratory's capabilities.
3
Optional items required to achieve specifications listed. Must use recommended options.
4
Example based on optional equipment (Keysight 6811B source and Yokogawa WT310E AC
standard) . Other options are available.
5
Assumes proper preparations are followed prior to storage, including draining the coolant
system.
1314-900 Revision BA
14 to +122 °F (-10 to +50 °C)
1-year Parts & Workmanship for AE-TEGAM-manufactured
components
Typical < 0.025 at f < 100 MHz, and <0.1 at f >= 100 MHz
Figure 1: Typical S
Required for routine calibration of the Calorimeter.
Required for routine calibration of the calorimeter.
Options available based on specific configurations.
For control of calorimeter and ancillary equipment.
Contact TEGAM for further details and options.
250W Precision Calorimeter
5
Linear Mag Performance